General
Protocol for Testing Systems that Treat Dental Unit Water
(Goal
= <
200 cfu/ml)
17 March, 2000
Return to Dental Unit Water
Menu
Research Question: Does the
unit, disinfectant, or unit-disinfectant combination maintain bacterial levels
in the dental unit water within the ADA goal of less than or equal to 200 colony
forming units per milliliter (cfu/ml) at the exit port of handpieces and air/water
syringes?
Methods
- Installation of dental unit waterline disinfection units
- Install disinfectant unit onto dental unit according to manufacturer
directions.
- The unit installation is to be performed by either by CRA
personnel, maintenance personnel of the office’s choice, or by the manufacturer.
- As per manufacturer’s directions, install only part of the
disinfecting unit (such an injection port). Installation by CRA personnel,
maintenance personnel of the office’s choice, or by the manufacturer.
- In office use of dental unit waterline disinfection unit, disinfectant,
or unit-disinfectant combination
- Dental Unit Waterline Disinfection Unit Only.
- Operate dental unit waterline disinfection unit according
to manufacturer’s directions. Operation to be conducted either by CRA staff
members or by office staff members with CRA supervision.
- Maintenance or servicing of the dental unit waterline disinfection
unit to be performed according to manufacturer’s directions, either by CRA
personnel, the office staff, maintenance personnel of the office’s choice,
or by the manufacturer.
- Disinfectant Only.
- The following options, per manufacturer’s directions, can
be employed to introduce the disinfectant into the dental unit waterlines:
- Disinfectant is added through an injection port.
- Disinfectant is added to the reservoir of an independent
water system that is either already in use by the office or has been
previously evaluated by CRA.
- After the recommended contact time, flush the disinfectant
from the lines, as per the manufacturer’s directions, by one of or by a
combination of the following methods:
- For colored disinfectants, flush until disinfectant is
no longer visible.
- For colorless disinfectants, flush until disinfectant
cannot be detected by recommended monitoring device (pH strips, etc.).
- Flush until the smell and/or feel of the disinfectant
is no longer detected.
- Flush for recommended time (usually 30 seconds to 1 min).
- Unit-Disinfectant Combination.
- Operate the dental unit waterline disinfection unit according
to manufacturer’s directions. Operation by either CRA staff members or by
office staff members with CRA supervision.
- Add the disinfectant recommended by the manufacturer. Addition
of the disinfectant performed by either CRA staff members or by office staff
members with CRA supervision.
- If disinfectant is automatically added to the dental unit
waterlines by the unit, monitor disinfectant concentration and made sure
the reservoir(s) are kept filled with disinfectant.
- If there is a recommended contact time, flush disinfectant
from the lines, as per the manufacturer’s directions, by one of or a combination
of the following methods:
- For colored disinfectants, flush until disinfectant is
no longer visible.
- For colorless disinfectants, flush until disinfectant
cannot be detected by recommended monitoring device (pH strips, etc.).
- Flush until the smell and/or feel of the disinfectant
is no longer detected.
- Flush for recommended time (usually 30 seconds to 1 min).
- Treat patients, per normal procedures
- Collection of Water Samples
- Attach sterile handpiece or air/water syringe tip to end
of line, then spray 5 - 50 ml of water into properly labeled sterile bottles.
Avoid touching the tips of handpieces on the bottle lip. Use of sterile
gloves is recommended.
- Send water samples to the lab either:
- Overnight in refrigerated pack or on ice if travel time will
be more than 30 minutes.
- At room temperature, if travel time will be less than 30
minutes.
- Once samples are received at the lab, refrigerate until microbial
evaluation. Evaluate for microbial load as soon as possible after receipt.
- Microbial Evaluation of Water Samples
- Controls.
- Negative control: sterile distilled water.
- Positive control: tap water sample.
- Neutralization controls: place duplicate 1-ml aliquots of
test water or sterile PSS ([Physiological Sterile Saline] for a neutralization
system control) into 9 ml of appropriate sterile neutralizer. Add a known
titer of Staphylococcus aureus. Allow 30 minute contact time and
then filter entire neutralization control (see step V.C.).
- Serially dilute water samples and positive control 1:10,
first in appropriate sterile neutralizer (see step V.G.c.) and then further
in PSS.
- Aseptically filter duplicate 1-ml aliquots through a 0.45-m
m filter (unless sample is a neutralization control, see V.A.c.).
- Place filter onto R2A (water samples) or TSA (S. aureus
controls) agar plates. Use plates that have been poured within the week
and refrigerated.
- Incubate TSA agar plates upside down for 18 – 24 hours at
37° C.
- Place the R2A plates upside down inside a plastic bag and
close with a twist tie (to prevent loss of moisture from the plates). Incubate
R2A agar plates for 7 days at room temperature (~
20 - 25oC).
- Count control colonies and calculate total cfu/ml of the
sample.
- Negative control is acceptable only if no colonies are found.
- Positive control is acceptable only if colonies are growing
on the R2A agar. The positive controls are compared over the testing period,
to verify that the titers are within the expected range and that the colonies
appear to be typical of what is normally found in city tap water samples.
- Neutralization controls are acceptable only if the S.
aureus titer is close to the titer added. If there is a significant
drop in the cfu/ml, then the neutralizer did not stop the action of the
disinfectant in the water sample, and another, more effective neutralizer
should be found and used.
- Count water sample colonies and calculate total cfu/ml of
the sample.
Return to Top
Return to Dental Unit Water
Menu